MRI Safety Information TRACOE vario and TRACOE silcosoft

REF 464

MR Safe

The TRACOE vario tracheostomy tube REF 464 is “MR Safe”.

REF 464

MRT-sicher

Die Tracheostomiekanüle REF 464 ist MRT Sicher.

REF 450, 451, 455

MR Unsafe

Keep the vario tracheostomy tubes REF 450,451 and 455 outside the MRI scanner room.

REF 450, 451, 455

Nicht MRT-sicher

Die vario Tracheostomiekanülen REF 450, 451 und 455 sind nicht MRT-sicher und sollten nicht im MRT Raum angewendet werden.

REF 460, 470, 461, 471

MR Conditional

Nonclinical testing has demonstrated the TRACOE vario tracheostomy tube is “MR Conditional”. A patient with this device can be safely scanned in an MR system meeting the following conditions:

In non-clinical testing, the image artifact, caused by the check valve, extends (radially) up to 107 mm from the check valve when imaged with a gradient echopulse sequence and a 1.5 T MR system, and up to 113 mm when imaged with a spin echo pulse sequence in a 3.0 T MR system. Therefore, it is recommended to tape the check valve to the patient’s skin away from the area of interest.

REF 460, 470, 461, 471

Bedingt MRT-sicher

Diese Person trägt eine TRACOE vario Tracheostomiekanüle und kann nur unter sehr spezifischen Bedingungen sicher im MRT gescannt werden. Das Scannen unter anderen Bedingungen kann zu einer Verletzung des Patienten führen. Vollständige Informationen zur MRT-Sicherheit finden Sie im Abschnitt MRT der TRACOE vario Gebrauchsanweisung, die mit dem Produkt mitgeliefert wird oder unter www.atosmedical.com/ifu erhältlich ist.

When used in MR Imaging:

Bei Anwendung unter MRT-Bildgebung:

REF 360, REF 361, REF 362, REF 363, REF 370, REF 371, REF 372, REF 373

MR Conditional

Non-clinical testing has demonstrated the TRACOE silcosoft tracheostomy tube is MR Conditional. A patient with this device can be safely scanned in an MR system meeting the following conditions:

In non-clinical testing, the image artifact caused by the tube extends (radially) up to 19 mm from the TRACOE silcosoft tracheostomy tube (non-cuffed) when imaged with a gradient echo pulse sequence and a 1.5T MR system, and up to 28 mm when imaged with a spin echo pulse sequence in a 3.0T MR system.

Tubes with H2O Cuff: In non-clinical testing, the image artifact, caused by the check valve, extends (radially) up to 107 mm from the check valve when imaged with a gradient echo pulse sequence and a 1.5T MR system, and up to 113 mm when imaged with a spin echo pulse sequence in a 3.0T MR system. Therefore, when MR imaging a tube with the H2O Cuff, it is recommended to tape the check valve to the patient’s skin away from the area of interest.

REF 360, REF 361, REF 362, REF 363, REF 370, REF 371, REF 372, REF 373

Bedingte MRT-Tauglichkeit

Nicht-klinische Untersuchungen haben ergeben, dass die TRACOE silcosoft Tracheostomiekanüle bedingt MRT-tauglich ist. Patienten mit diesem Produkt können sicher mit einem MRT untersucht werden, sofern folgende Bedingungen erfüllt sind:

In nicht-klinischen Untersuchungen erstreckte sich der durch die Kanüle verursachte Bildartefakt (radial) über bis zu 19 mm ab der TRACOE silcosoft Tracheostomiekanüle (ohne Cuff) bei einer Aufnahme mit Gradient-Echo-Pulssequenz in einem 1,5 T MR-System und bis zu 28 mm bei einer Aufnahme mit Spin-Echo-Pulssequenz in einem 3,0 T MR-System.

Kanülen mit H2O Cuff: In nicht-klinischen Untersuchungen erstreckte sich der durch das Absperrventil verursachte Bildartefakt (radial) über bis zu 107 mm ab dem Absperrventil bei einer Aufnahme mit Gradient-Echo-Pulssequenz in einem 1,5 T MR-System und bis zu 113 mm bei einer Aufnahme mit Spin-Echo-Pulssequenz in einem 3,0 T MR-System. Daher wird empfohlen, bei MRT-Aufnahmen mit einer Kanüle mit H2O Cuff das Absperrventil fern der zu untersuchenden Region auf der Haut des Patienten zu fixieren.

When used in MR Imaging:

Bei Anwendung unter MRT-Bildgebung: