MRI Safety Information TRACOE twist

REF 301, REF 302, REF 306, REF 888-306

MR Conditional

Nonclinical testing has demonstrated the TRACOE twist tracheostomy tubes REF 301, REF 302, REF 306, REF 888-306 are “MR Conditional”. A patient with this device can be safely scanned in an MR system meeting the following conditions:

In non-clinical testing, the image artifact, caused by the check valve, extends (radially) up to 107 mm from the check valve when imaged with a gradient echopulse sequence and a 1.5 T MR system, and up to 113 mm when imaged with a spin echo pulse sequence in a 3.0 T MR system. Therefore, it is recommended to tape the check valve to the patient’s skin away from the area of interest.

REF 301, REF 302, REF 306, REF 888-306

Bedingt MRT-sicher

Diese Person trägt eine TRACOE twist Tracheostomiekanüle REF 301, REF 302, REF 306, REF 888-306 und kann nur unter sehr spezifischen Bedingungen sicher im MRT gescannt werden. Das Scannen unter anderen Bedingungen kann zu einer Verletzung des Patienten führen. Vollständige Informationen zur MRT-Sicherheit finden Sie im Abschnitt MRT der TRACOE twist Gebrauchsanweisung, die mit dem Produkt mitgeliefert wird oder unter www.atosmedical.com/ifu erhältlich ist.

REF 303, REF 304, REF 305

MR Safe

The TRACOE tracheostomy tubes REF 303, REF 304, REF 305 are "MR Safe".

REF 303, REF 304, REF 305

MRT-sicher

Die Tracheostomiekanülen REF 303, REF 304 und REF 305 sind MRT Sicher.

When used in MR Imaging:

Bei Anwendung unter MRT-Bildgebung: